Repairing Guides

how to repair pressure gauge

by Judy Cremin Published 2 years ago Updated 2 years ago
image

5 Techniques to Adjust or Repair a Pressure Gauge

  • Opening a vent to release pressure build-up
  • Opening the Glass (or plastic cover) then removing the needle.
  • Rotating the screw located at the face of the scale.
  • Adjusting the knob at the lower side of the pressure gauge
  • Adjusting the screw located at the gauge needle pointer itself.

Full Answer

How do you fix a pressure gauge that keeps pointing off?

The gauge may constantly point off of zero when there’s no pressure in the line or the gauge is consistently reading 5% above or below the range it really should be reading based on your other instrumentation. One solution to this problem is to utilize the adjustable pointer on the pressure gauge assuming your gauge has one.

Why do pressure gauges have adjustable pointers?

Most industrial pressure gauges will have an adjustable pointer because gauges will commonly fall out of calibration out in the field due to vibration, pulses, or just general use. Gauges that don’t typically have them are smaller utility pressure gauges or liquid filled gauges.

What does an industrial pressure gauge look like?

It would look something like this: Most industrial pressure gauges will have an adjustable pointer because gauges will commonly fall out of calibration out in the field due to vibration, pulses, or just general use. Gauges that don’t typically have them are smaller utility pressure gauges or liquid filled gauges.

How do you refurbish a gauge?

Step 1 - Old gauges are received for refurbishment. Step 2 - Gauge is completely disassembled and evaluated. Step 3 - Interior and exterior surfaces of the case and ring are cleaned and repainted at our paint station. Mounting threads are also cleaned.

image

Can a pressure gauge be repaired?

At Gage-it we're experts in the repair and refurbishment of pressure gauges and related accessories. With over 45 years experience in the gauge repair industry, Gage-it can restore just about any gauge regardless of condition to its new or original condition.

Why is my pressure gauge not working?

Pressure gauge failure can be attributed to one or more of these eight reasons: mechanical vibration, pulsation, extreme temperature, pressure spikes, overpressure, corrosion, clogging, and mishandling/abuse.

How do you fix a pressure gauge needle?

1:343:09How To Burp A Pressure Gauge - YouTubeYouTubeStart of suggested clipEnd of suggested clipSo what we're going to do is are gonna take a flathead screwdriver. Or a long finger now if you haveMoreSo what we're going to do is are gonna take a flathead screwdriver. Or a long finger now if you have it and on top there's a fill plug here.

How do you fix a broken gauge?

0:013:46Broken Gauge? FREE FIX! WILL IT WORK? - YouTubeYouTubeStart of suggested clipEnd of suggested clipAnd what you do is if your gauge is broken like this you can see the needle it's stuck below itsMoreAnd what you do is if your gauge is broken like this you can see the needle it's stuck below its little peg there. So I saw this trick what you do is you get a magnet.

How do you test a pressure gauge?

To perform the test:Isolate the pressure gauge from the process using valves, or by removing the gauge from the process.Connect the gauge to the calibrator or reference gauge. ... For hydraulic pressure gauges it's important prime the system. ... When generating pressure allow a few moments for the measurement to stabilize.More items...

How do you clean a pressure gauge?

Thermocouple Tube CleaningThe first step for effectively cleaning a digital pressure gauge thermocouple tube is to remove it from the attached cable. ... Using a funnel, pour roughly a tablespoon of solvent into the sensor's stem. ... The solvent needs time to work, so allow it to sit for about an hour.More items...

How do I know if my water pressure gauge is broken?

Pressure gauge problems are one of the easier well issues to diagnose. If the PSI gauge doesn't ever move, then it is most likely broken. Another warning sign is a pressure reading when the system has no water in it. The pressure gauge is a tool that allows us to see how the system is operating.

How does a pressure gauge work?

They consist of a semicircular and flat tube of metal, fixed at one end and attached to a sensitive lever mechanism at the other. As pressure increases inside the tube, the force of the fluid attempts to straighten out the curved tube.

Why does my pressure gauge have water in it?

To put it simply, liquid filling a pressure gauge is strictly used as an inexpensive way to steady the needle and lubricate the internals of a gauge during applications with vibration. The most common fill fluids used are Glycerin and Silicone.

Can you fix a broken gas gauge?

Check your owner's manual to determine where the fuse is located for the gas gauge. Once located open the fuel box and check all the fuses inside. If any are damaged or completely broken, replace them. If this doesn't seem to be an issue, you're going to move onto the next check, the sending unit.

How do you fix a cluster gauge?

1:0019:00How to repair GM Instrument Cluster | Speedometer | Gauge | StepperYouTubeStart of suggested clipEnd of suggested clipAnd it has little tabs in it you have to carefully undo without breaking them there's a fewMoreAnd it has little tabs in it you have to carefully undo without breaking them there's a few different pieces you just got to go around and and fry in there and undo these little tabs.

Can I fix my own gas gauge?

If you are looking to do a gas gauge repair by yourself, it can be done. In older vehicles it is much easier to diagnose gas gauge problems because they consist of manual gauges. When there are issues with a manual gauge you may see that the lever goes below the empty meter on the gas panel.

How do I know if my pressure gauge is bad?

Symptoms of a bad oil pressure gaugeOil pressure gauge not working: The causes for this range from a faulty gauge to the need for an oil change. ... Oil pressure gauge reading too low, generally below 15 to 20 PSI while idling. ... Oil pressure gauge reading too high, or over 80 PSI while driving, especially at higher RPMs.

How do I know if my water pressure gauge is broken?

Pressure gauge problems are one of the easier well issues to diagnose. If the PSI gauge doesn't ever move, then it is most likely broken. Another warning sign is a pressure reading when the system has no water in it. The pressure gauge is a tool that allows us to see how the system is operating.

How long do pressure gauges last?

According to NFPA 25, Standard for the Inspection, Testing, and Maintenance of Water-Based Fire Protection Systems, gauges shall be replaced every 5 years or tested every 5 years by comparison with a calibrated gauge.

How do you reset a liquid filled pressure gauge?

0:502:28How to Reset a Liquid Filled Pressure Gauge - YouTubeYouTubeStart of suggested clipEnd of suggested clipAll you have to do is fit a fingernail into the cap to release the pressure and the pointer quicklyMoreAll you have to do is fit a fingernail into the cap to release the pressure and the pointer quickly returns to normal no muss no fuss. You can actually use this process with any of the gages.

1. There are differences between diaphragm pressure gauges and standard pressure gauges in terms of pressure measurement principle and structure

The pressure measuring principle of the standard pressure gauge is to be calculated into the pressure measuring spring. The spring tube is under pressure, and it will produce elastic displacement and drive the connecting rod and the central gear deflection. The pointer is mounted on the main gear shaft to show the pressure measurement value.

2. Analysis of the causes of faults in the use of diaphragm pressure gauges

2.1 The diaphragm pressure gauge is distributed in the TEC boundary zone, and the failure rate

3. Diaphragm pressure gauge servicing

We have overhauled more than one diaphragm pressure gauge, and we found that most of the failures are caused by the loss of silicone oil in the spring tube. So it can be said that the spring tube filled with silicone oil solves the key problem of diaphragm pressure gauge maintenance. But to the spring tube filled with silicone oil is not easy.

4. Calibration of diaphragm pressure gauges

Our diaphragm pressure gauges are supplied from abroad with the equipment and have a standard 1/2 NPT thread to connect the pressure gauge head. In addition, each diaphragm pressure gauge is provided with a connection flange. Therefore, we want to calibrate the diaphragm pressure gauge.

5. Conclusion

The diaphragm pressure gauges are widely used in oil transmission stations for the primary measurement of operating pressure. Still, they also have the problem of being damaged and expensive to purchase new gauges.

Fault 2: The pointer of the pressure gauge is bouncing or sluggish

1. There is friction between the pointer and the surface glass or dial. You can correct the pointer, thicken the washer under the glass or twist the pointer shaft hole-some.

Fault 3: Scale deviation

1. At which scale you find out of tolerance, you stop at that scale. Check the coordination of various parts on the scale point, whether the transmission shaft hole is blocked; whether the connecting rod is flexible; whether the tooth meshing point is damaged or foreign objects are eliminated.

Fault 4 Uneven pressure indication error

1. The deformation of the spring tube fails, and the displacement and the pressure are not proportional to each other. The spring tube needs to be replaced.

What is a liquid filled pressure gauge?

Liquid filled pressure gauges come with a variety of different stoppers that keep the liquid inside. It may have a rubber stopper, a rubber stopper with a brass pin, or a closed pressure release tube. Although they look a bit different, the key to resetting them is the same. You want to relieve pressure by venting the cap.

Why is my ship gauge inaccurate?

During shipping, the gauge may have experienced high temperature and altitude changes. This causes the internal pressure to build inside the gauge, thus resulting in inaccurate readings. Don’t worry. The gauge can be re-calibrated.

How to fix a rubber stopper?

Rubber Stopper With a Brass Pin. Gauges with a brass pin are even easier to fix. Simply pull the pin to release the pressure. It should set itself back to zero. After that, you can push the pin back in place. Other Tips. Once you have vented your liquid filled pressure gauge, it is unlikely that it will happen again.

Should a pressure gauge be installed vertically?

That is, as long as the gauge isn’t exposed to high temperature or pressure drops. (This is not often the case.) The gauge should be installed vertically, especially if it’s capped by a pressure relief tube at the top. This prevents the glycerin from leaking and makes it easier to vent the gauge later on, if needed.

image

Analysis of The Causes of Faults in The Use of Diaphragm Pressure Gauges.

Image
2.1 The diaphragm pressure gauge is distributed in the TEC boundary zone, and the failure rate 2.2 Analysis of the causes of several common faults. 2.2.1 When there is pressure, its pointer does not indicate. This failure is usually because solid vibrations loosen the connection between the pointer and the central gear shaft, a…
See more on bcstgroup.com

Diaphragm Pressure Gauge Servicing

  • We have overhauled more than one diaphragm pressure gauge, and we found that most of the failures are caused by the loss of silicone oil in the spring tube. So it can be said that the spring tube filled with silicone oil solves the key problem of diaphragm pressure gauge maintenance. But to the spring tube filled with silicone oil is not easy. In general, the manufacturer has a special sil…
See more on bcstgroup.com

Calibration of Diaphragm Pressure Gauges

  • 4.1 Production of special calibration tools
    Our diaphragm pressure gauges are supplied from abroad with the equipment and have a standard 1/2 NPT thread to connect the pressure gauge head. In addition, each diaphragm pressure gauge is provided with a connection flange. Therefore, we want to calibrate the diaphra…
  • 4.2 Diaphragm pressure gauge calibration procedure
    4.2.1 First, we should remove the head section of the diaphragm pressure gauge from the flange. Using the special connector shown in Figure 2, connect the pressure gauge to the calibrator and adjust it to pass the calibration according to the standard pressure gauge calibration method. 4.…
See more on bcstgroup.com

Conclusion

  • The diaphragm pressure gauges are widely used in oil transmission stations for the primary measurement of operating pressure. Still, they also have the problem of being damaged and expensive to purchase new gauges. As a result, we have to repair faulty or damaged diaphragm pressure gauges ourselves to reduce the purchase of new meters and reduce production and op…
See more on bcstgroup.com